About the product
Four operating modes : CMV, A/CMV, SIMV, SPONT
Lightweight, durable and simple to use
Ventilation volume and ventilation frequency can be adjusted in succession and separately
LCD graphical display to indicate airway pressure, alarm limits, tidal volume, mode
Automatic pressure relief valve for airway pressure more than 6KPa
Air-oxygen blender prolongs the operating time during long-time transportation
Audiovisual low input pressure alarm and over input pressure protection
Specification
VENTILATOR PARAMETERS
RANGE
Tidal Volume (at C mode)
50~1200 ml
Flow (at A/C, SIMV mode)
Low, Med, High
Breath Rate
5-60 bpm, 1~12 bpm for SIMV
%02
45%~90%
Spont Flow
0~5 L/Min
Sigh
1 per 0~100 breath cycle
Inspiratory time
0.5~3 s
Trigger Level
-10~0 hpa
PIP
0~60 hpa
Silence
60s
ALARMS AND INDICATORS
High Pressure Alarm Limit
20~60hpa (to be set)
Low Pressure Alarm Limit
2~20hpa (to be set)
Alarm Indicator
Flashing Red Indicator
Over Pressure Relief
Lightweight, durable and simple to use
Ventilation volume and ventilation frequency can be adjusted in succession and separately
LCD graphical display to indicate airway pressure, alarm limits, tidal volume, mode
Automatic pressure relief valve for airway pressure more than 6KPa
Air-oxygen blender prolongs the operating time during long-time transportation
Audiovisual low input pressure alarm and over input pressure protection
Specification
VENTILATOR PARAMETERS
RANGE
Tidal Volume (at C mode)
50~1200 ml
Flow (at A/C, SIMV mode)
Low, Med, High
Breath Rate
5-60 bpm, 1~12 bpm for SIMV
%02
45%~90%
Spont Flow
0~5 L/Min
Sigh
1 per 0~100 breath cycle
Inspiratory time
0.5~3 s
Trigger Level
-10~0 hpa
PIP
0~60 hpa
Silence
60s
ALARMS AND INDICATORS
High Pressure Alarm Limit
20~60hpa (to be set)
Low Pressure Alarm Limit
2~20hpa (to be set)
Alarm Indicator
Flashing Red Indicator
Over Pressure Relief
Contact with supplier
Company